Implantable Prosthetic Heart Valves


Introduction        Clinical Background        Design Considerations        Mechanical Valves        Tissue-based Valves        Future Developments        References

Design Considerations

Although the primary concern is to mimic the function of the native heart valve, there are secondary considerations that need to be addressed as well.  The secondary considerations revolve around minimizing unwanted side effects that result from the design of the prosthetic valve.

  1. Biocompatibility – Materials used must not provoke an immune response.  This means presenting infection as well as thrombosis or calcification that may impede the operation of the new valve.
  2. Durability – Heart valves are expected to work continuously for several years and are implanted though a major invasive surgery.  For these two reasons, we cannot afford to have valves fail prematurely (see figure 4).
  3. Manufacturability – The construction of the valve should be feasible and consistent.  This requires us to consider materials that have high yield rates.  The design must be able to be constructed reasonably (see next consideration).
  4. Cost – The cost of the valve must be reasonable and competitive with similar products.  Even if all of the above design considerations were perfectly met, there would be resistance in the marketplace if the cost of the valve were too high.

Figure 4: An accelerated heart vavle wear tester.  To ensure that heart valve designs can withstand the constant forces and motion, machines such as this are used to simulate the number of cardiac cycles the valves are expected to endure through their expected lifetime.  Thus, several years worth of valve activity are compressed into a few weeks or months.

Previous                    BME 240 Home                    Next

Page design and content by Vincent Nguyen-Hoai